F-950 Three Gas Analyzer



The F-950 Three Gas Analyzer measures levels of ethylene, CO₂ and O₂ to maintain optimal produce quality at every phase of the supply chain. The F-950’s versatile design allows it to be scaled to many environments, from cold storage to transportation containers. Simple to operate and weighing less than a kilogram, the F-950 uses an electrochemical cell to measure ethylene between 0–200 ppm. The F-950 is ideal for measuring ethylene production across a wide range of fruit types.

Key Applications:

  • Verification of ethylene mitigation systems for increasing commodity shelf life and storage longevity.
  • Monitoring controlled atmosphere facilities to ensure the proper balance of gasses throughout operations.
  • Modified Atmosphere Packaging QA in packing lines, laboratories, import/export surveys, distribution centers, and retail centers.
Precision citrus degreening and ripening control on various commodities.

Parameter

Specification

Sampling Flow Rate

70 mL/min

Total Sampling Time

10 seconds

Data Saving

Automated, 1-second intervals

Display

Backlit visible transflective LCD

Operating Environment

0–50°C, 0–90% RH (non-condensing)

Operating Atmospheric Pressure

500–1200 mbar

Dimensions

18 × 13.5 × 5.5 cm

Weight

0.95 kg

Enclosure

Powder-coated aluminum

Battery Life

8.5 hours

Power Source

Removable rechargeable Li-ion battery

PC Interface

USB, Wi-Fi, SD card, Bluetooth

Data Recorded

C₂H₄, CO₂ & O₂ concentrations, date, time, RH, GPS location

Parameter

C₂H₄ – Trigger

CO₂ – Trigger

O₂ – Trigger

C₂H₄ – Continuous

CO₂ – Continuous

O₂ – Continuous

Sensor Type

Electrochemical

Infrared, Pyroelectric

Electrochemical

Electrochemical

Infrared, Pyroelectric

Electrochemical

Range

0–200 ppm

0–100%

0–100%

0–200 ppm

0–100%

0–100%

Resolution

0.1 ppm

0.01%

0.10%

0.1 ppm

0.01%

0.10%

Lower Detection Limit

0.15 ppm

0.50%

0.30%

0.2 ppm

0.01%

0.10%

Sampling Volume

35 mL

12 mL

12 mL

Response Time

35 s

12 s

12 s

80 s

6 s

12 s

Offset Calibration

Weekly

Weekly

Weekly

Weekly

Weekly

Weekly

Sensor Replaceable

Yes

No

Yes

Yes

No

Yes

Accuracy (Absolute)

±2 ppm

±0.50%

±0.30%

±0.2 ppm

±0.01%

±0.10%

Accuracy (Relative)

±5%

±3%

±2%

±5%

±3%

±2%

Span Calibration

3 months

12 months

6 months

3 months

12 months

6 months

Lifetime

2 years

>15 years

2 years

2 years

>15 years

2 years